Several years ago I started working with a patient who has a very significant hearing loss with an unusual configuration. He chose hearing aids anyway and had an unexpectedly great outcome, understanding family, friends, TV dialogue and even conversation at a restaurant table. After four years he upgraded his technology and heard even better.
Shortly after that upgrade, things went horribly wrong, to the point where he was scheduled for cochlear implant surgery he didn't actually need. What happened to him should never happen, yet it happens almost all of the time in the hearing aid industry, and almost nobody talks about it.
A Cochlear Implant Referral That Didn't Add Up
My patient is a snowbird who spends about six months of the year in Phoenix and six months up north. I saw him in the spring before he left, and he was doing great with no complaints. I had told him he needed a provider he could see when he wasn't in the greater Phoenix area.
While he was up north he started having trouble hearing, so he went to that clinic. They gave him an updated hearing test and found no change in his hearing at all. They did a listening check on his hearing aids, everything sounded fine to them, and they turned up the volume and sent him home to try it for a few days. When the problems continued, they told him there was nothing else they could do and referred him for a cochlear implant candidacy evaluation.
The implant center tested him with his hearing aids, decided he was a candidate, and put him on the schedule for surgery. We had talked about implants before as a future option, but he started having second thoughts and wanted my final stamp of approval first.
When he told me the story, it didn't make sense. He brought in his new audiogram, and it was virtually identical to the one I had done about six months earlier, when he was doing great.
What a Test Box Found in Minutes
I cleaned his hearing aids and did my own listening check, and I couldn't tell whether they sounded right or not. With a hearing loss as significant and unusual as his, a listening check just doesn't tell you much. So I ran an objective diagnostic test on his hearing aids, something none of the other audiologists had done.
Within a few minutes I knew the problem. Both hearing aids were mechanically malfunctioning. They still produced sound, but the level was lower than it needed to be, and the sound was badly distorted and noisy. You could argue he was hearing worse with them in his ears than he would have with them out.
We fit him right away with loaner hearing aids, using the same programming he had before he left. When I put them in his ears, he said, "Oh my god, my hearing's back." He was about to have cochlear implant surgery because no one ran a diagnostic measure on his hearing aids, and every bit of it could have been prevented with test box measures. If you want to know when an implant referral really does make sense, I cover that in my video on cochlear implant referral criteria.
What Hearing Aid Test Box Measures Are
Hearing instrument test box measures, also called electroacoustic analysis, are a series of diagnostic tests that objectively show whether your hearing aids are mechanically working the way they should. They mean your provider doesn't have to rely on a simple listening check.
It's the equivalent of an oncologist ordering an MRI to look for a tumor instead of relying only on your reported symptoms. A cancer doctor who never ordered imaging would likely lose their license for negligence.
There are two separate ideas here. A hearing aid that is programmed correctly is different from a hearing aid that is functioning correctly. You can have perfect programming, and it won't matter if the device fails diagnostic testing. That's why test box measures sit alongside Real Ear Measurement as one of the audiologic best practices your provider should be performing.
How the test works
After cleaning your hearing aids, we put them into a specific diagnostic test setting and place them in a calibrated sound chamber. We play precisely controlled sounds, measure exactly what the hearing aid amplifies, and compare the results to the manufacturer's specifications for that model.
A hearing aid that is within specs shows smooth output and gain curves, with every value inside the tolerance ranges on the manufacturer's spec sheet. A hearing aid that is out of specs usually shows curves that sit too low or aren't smooth, or values outside those tolerance ranges. When I compared my patient's measurements from before his summer trip with the ones I took after he came back, the difference was dramatic. The first set looked great. The second had jagged curves, output and gain numbers that were far too low, and distortion and noise numbers that were far too high.
Why Listening Checks Miss So Much
I've been doing this for 14 years, and my listening checks still aren't good enough. They are entirely subjective. It typically takes 50 to 100% distortion of a sound before a hearing care professional will detect it by listening, and by then it's destroying your performance. A hearing aid can be out of spec with only about 4% distortion, which is already enough to hurt your hearing.
Research shows that roughly 15 to 20% of hearing aids aren't meeting manufacturer specifications. In my own clinic, nearly 50% of new patients who arrive already wearing hearing aids have at least one device that fails diagnostic testing.
That's why, at Applied Hearing Solutions, we run a test box measure on every patient's hearing aids every year, even when they have no complaints, so we can catch devices drifting out of spec before it becomes a big problem. Test box measures should also be done every time you're fit with new hearing aids and every time your hearing aids come back from a manufacturer repair. My patient's story shows why. His first repair came back with one hearing aid still failing, and we had to send it back again before both met specifications.
How to Check Whether Yours Have Been Tested
Ask your provider this question: "Have you performed test box measures on my hearing aids to verify that they are within manufacturer specifications?"
If the answer is yes, ask to see the results. They can be on paper or on a screen. Look for values such as:
- OSPL90 max (or peak) and OSPL90 HFA
- Full-on gain (FOG) max or peak
- Total harmonic distortion (THD)
- Equivalent input noise (EIN)
If they can't show you these results, I'd treat that as a red flag and ask them to run the test again so you can see it. This matters even more if your hearing aids are a couple of years old. For more questions worth bringing to your appointment, see my list of 5 questions to ask your provider.
Your hearing aids don't have to fail as badly as my patient's did to hold you back. Even a device that is slightly out of spec keeps you from hearing your best, and without a test box measure, neither you nor your provider will know. My patient is incredibly happy with his hearing aids again, and he skipped a surgery he didn't want. He also knows that if he ever has hearing aid problems at another clinic, he should insist on a test box measure before anyone talks to him about surgery.
Find a Provider Who Runs Test Box Measures
Finding a hearing care professional who performs test box measures is even harder than finding one who does Real Ear Measurement. HearingUp Network members are committed to comprehensive audiologic best practices, including test box measures on every new and repaired hearing aid. You can see the full list on the HearingUp Best Practices checklist.